CAEN N93B

说明书 N93B

DUAL TIMER - mod 2255B

该插件由两组完全相同的触发脉冲产生器组成。可用来调节 NIM 及 ECL 信号的宽度从 50 ns 到 10 s。

The END-MARKER, a short output pulse at the end of the tiiming cycle, can be uesd for delayed triggering of other units or, if fed back into the START input, it provides a self-running rate generator that can be started and stopped at will.

It is possible to disable start inputs by feeding a NIM signal to the "Veto" input.

输入

  • START INPUT
    • NIM and ECL,leading edge sensitive.
    • Minimum width 5 ns.
    • Momentary switch for single cycle operation.
  • VETO INPUT
    • NIM input
    • NUM ture level at VETO input desables the START input (within +-2ns of START leading edge)
  • RESET INPUT
    • NIM input
    • Minimum width: 7 ns (15 ns for recursive operation)
    • It can be applied at any point of the time cycle.
    • Produces END-MARKER

输出

  • OUTPUTS
    • Two normal independent,one complementary NIM output and one ECL.
  • WIDTH
    • 50 ns to 10 s in 9 decade steps, with a potentiometer and locking dial for fine adjustment
    • Accuracy +-10% of full scale, with adequate range overlap.
    • Temperature coefficient: 10^{-3}/^{0}C of setting
    • Switch postion <<无穷>> for bistable opreation
  • DEAD TIME
    • Shorter than the cycle time
    • The timer can be triggered well before the ehd of the timing cucle.
  • RISE/FALL TIME
    • <= 2 ns
  • INPUT-OUTPUT DELAY
    • Delay from Start to leading edge of output,or from Reset to trailing edge,is ~13 ns
  • END MARKER
    • NIM and ECL output, 15 ns wide output pulse leading edge is coincident with the trailing edge of output within 2 ns.
    • Can be connected to Start input for free running opreation (sequence activated by the START switch)
  • LED
    • Flashes for 0.1 s or output width, whichever is longer.
    • At high rates the lamp has its own flashing rate.

results matching ""